Effective radiation doses in neck computed tomography scans

  • Khalid Alzimami
  • , Layal Jambi
  • , Essam Mattar
  • , Ahmed Alenezi
  • , Abdulrahman Alfuraih
  • , Hassan Salah
  • , Mohammad Rabbaa
  • , Mohammad Abuljoud
  • , Khaled Alsafi
  • , Abdullah Alsubaie
  • , Abdelmoneim Sulieman
  • , David A. Bradley
  • King Saud University
  • Inaya Medical College
  • Sudan University of Science and Technology
  • Riyadh Care Hospital
  • King Abdulaziz University
  • Taif University
  • Sunway University
  • University of Surrey

Research output: Contribution to journal › Article › peer-review

8 Scopus citations

Abstract

A significant source of ionizing radiation for medical use is computed tomography (CT). According to estimates, more than 84 million CT scans are carried out yearly in the US. Because the neck area contains several organs and tissues that are very sensitive to ionizing radiation, such as the thyroid gland, CT for the neck region is commonly done for a variety of clinical causes. This study's goal is to assess the patient doses to radiosensitive organs and effective doses received during the CT neck surgery in the pastAt Riyadh Care Hospital, a multislice CT (MSCT) system (Siemens Somatom Sensation, 64 slices/detector) was used to perform improved CT treatments on a total of 54 patients. Patient ages were 41.5–15.5 years old on average, with a wide variation (21–73). Using computer software based on Monte Carlo simulation, organ and effective dosages were assessed. The International Commission of Radiological Protection (ICRP) risk variables were used to calculate the patient's radiogenic risk for a particular organ. The patients' dose length product (DLP, mGy.cm) for CT neck examination were 993.3 ± 539.5 (267.3–2441.7). The mean and range of volume CT dose per slice (CTDIvol (mGy)) were 1.5 ± 3.9 (3.91–22.55). The overall effective dose (mSv) per procedure is ranged from 1.3 to 11.7 with a mean value 2.8 mSv. Patients' CT neck procedures have been repeated the procedure up to 7 times during the treatment course. The risk of malignancy is 1x10-6 CT procedure. The study showed that patients received wide discrepancies in patients' doses up to 9 times. Optimization of image acquisition protocol and scan length is necessary to reduce the unnecessary risk.

Original languageEnglish
Article number110340
JournalRadiation Physics and Chemistry
Volume200
DOIs
StatePublished - Nov 2022

Fingerprint

Dive into the research topics of 'Effective radiation doses in neck computed tomography scans'. Together they form a unique fingerprint.

Cite this