T1 - Abdominal aorta measurements by a handheld ultrasound device compared with a conventional cart-based ultrasound machine
AU - Alfuraih, Abdulrahman M.
AU - Alrashed, Abdulaziz I.
AU - Almazyad, Saleh O.
AU - Alsaadi, Mohammed J.
N1 - Publisher Copyright:
Copyright © 2021, Annals of Saudi Medicine, Saudi Arabia.
PY - 2021/11
Y1 - 2021/11
N2 - BACKGROUND: Ultraportable or pocket handheld ultrasound devices (HUD) may be useful for large-scale abdominal aortic aneurysm screening. However, the reproducibility of measurements has not been compared with conventional cart-based ultrasound machines. OBJECTIVES: Investigate the intra- and inter-operator reproducibility of a HUD compared with a conventional ultrasound machine for aortic screening. ESIGN: Analytical, cross-sectional. SETTING: Ultrasound department at a large tertiary care hospital in Riyadh. PATIENTS AND METHODS: Eligible male participants aged ≥60 years were invited to participate upon arriving for a non-vascular ultrasound appointment. Three repeated anteroposterior measurements of the transverse aorta were made at the proximal and distal locations for each machine before repeating the measurements on a subset of participants by a second blinded operator. Intraclass correlation coefficients (ICC) and the Bland-Altman method were used to analyze reproducibility. MAIN OUTCOME MEASURE: Inter-system and intra- and inter-operator ICCs. SAMPLE SIZE: 114 males with repeated measurements by second operator on a subset of 35 participants. RESULTS: The median age (interquartile range) of participants was 68 years (62-74 years). The intra- and inter-operator ICCs were all >0.800 showing almost perfect agreement except for the inter-operator reproducibility at the proximal location using a conventional machine (ICC= 0.583, P=.007) and the Butterfly device (ICC=0.467, P=.037). The intersystem ICCs (95% CI) were 0.818 (0.736-0.874) and 0.879 (0.799-0.924) at the proximal and distal locations, respectively. The mean difference in aortic measurement between the ultrasound systems was 0.3 mm (1.7%) in the proximal location and 0.6 mm (3.6%) in the distal location. In total, >91% of the difference in measurements between the machines was <3 mm. The mean scanning time was 4:16 minutes for the conventional system and 3:53 minutes for the HUD (P=.34). CONCLUSIONS: Abdominal aortic screening using a HUD was feasible and reliable compared with a conventional ultrasound machine. A pocket HUD should be considered for large-scale screening. LIMITATIONS: No cases of abdominal aortic aneurysm in the sample and lack of blinding. CONFLICT OF INTEREST: None.
AB - BACKGROUND: Ultraportable or pocket handheld ultrasound devices (HUD) may be useful for large-scale abdominal aortic aneurysm screening. However, the reproducibility of measurements has not been compared with conventional cart-based ultrasound machines. OBJECTIVES: Investigate the intra- and inter-operator reproducibility of a HUD compared with a conventional ultrasound machine for aortic screening. ESIGN: Analytical, cross-sectional. SETTING: Ultrasound department at a large tertiary care hospital in Riyadh. PATIENTS AND METHODS: Eligible male participants aged ≥60 years were invited to participate upon arriving for a non-vascular ultrasound appointment. Three repeated anteroposterior measurements of the transverse aorta were made at the proximal and distal locations for each machine before repeating the measurements on a subset of participants by a second blinded operator. Intraclass correlation coefficients (ICC) and the Bland-Altman method were used to analyze reproducibility. MAIN OUTCOME MEASURE: Inter-system and intra- and inter-operator ICCs. SAMPLE SIZE: 114 males with repeated measurements by second operator on a subset of 35 participants. RESULTS: The median age (interquartile range) of participants was 68 years (62-74 years). The intra- and inter-operator ICCs were all >0.800 showing almost perfect agreement except for the inter-operator reproducibility at the proximal location using a conventional machine (ICC= 0.583, P=.007) and the Butterfly device (ICC=0.467, P=.037). The intersystem ICCs (95% CI) were 0.818 (0.736-0.874) and 0.879 (0.799-0.924) at the proximal and distal locations, respectively. The mean difference in aortic measurement between the ultrasound systems was 0.3 mm (1.7%) in the proximal location and 0.6 mm (3.6%) in the distal location. In total, >91% of the difference in measurements between the machines was <3 mm. The mean scanning time was 4:16 minutes for the conventional system and 3:53 minutes for the HUD (P=.34). CONCLUSIONS: Abdominal aortic screening using a HUD was feasible and reliable compared with a conventional ultrasound machine. A pocket HUD should be considered for large-scale screening. LIMITATIONS: No cases of abdominal aortic aneurysm in the sample and lack of blinding. CONFLICT OF INTEREST: None.
UR - http://www.scopus.com/inward/record.url?scp=85120954993&partnerID=8YFLogxK
U2 - 10.5144/0256-4947.2021.376
DO - 10.5144/0256-4947.2021.376
M3 - Article
C2 - 34878929
AN - SCOPUS:85120954993
SN - 0256-4947
VL - 41
SP - 376
EP - 382
JO - Annals of Saudi Medicine
JF - Annals of Saudi Medicine
IS - 6
ER -