Abstract
This study aimed to assess patient entrance surface air kerma (ESAK) during chest and abdominal X-ray procedures in screen film radiography (SFR) and computed radiography (CR) to establish dose reference levels. Patients' doses were measured in five hospitals for a total of 196 patients. ESAK was calculated from exposure parameters using DosCal software. The X-ray tube output (mGy mAs-1), accuracy of exposure factors, linearity and reproducibility were measured using an Unfors Xi dosimeter. The overall mean and range of ESAK during chest X-ray were 0.6 ± 0.3 (0.1-1.3) mGy, while for abdominal X-rays they were 4.0 ± 3.2 (1.3-9.2) mGy. Hospital with a CR system was found to use relatively higher doses. Dose values for abdominal X-ray procedures were comparable with previous studies. The dose for chest X-ray procedure was higher by a factor of 2-3 compared with the current international reference levels.
| Original language | English |
|---|---|
| Pages (from-to) | 102-106 |
| Number of pages | 5 |
| Journal | Radiation Protection Dosimetry |
| Volume | 165 |
| Issue number | 1-4 |
| DOIs | |
| State | Published - 1 Jul 2015 |