Radiation dose levels for conventional chestand abdominal X-ray procedures in elected hospitals in Sudan

  • E. Babikir
  • , Hussein A. Hasan
  • , A. Abdelrazig
  • , M. A. Alkhorayef
  • , E. Manssor
  • , A. Sulieman

Research output: Contribution to journalArticlepeer-review

10 Scopus citations

Abstract

This study aimed to assess patient entrance surface air kerma (ESAK) during chest and abdominal X-ray procedures in screen film radiography (SFR) and computed radiography (CR) to establish dose reference levels. Patients' doses were measured in five hospitals for a total of 196 patients. ESAK was calculated from exposure parameters using DosCal software. The X-ray tube output (mGy mAs-1), accuracy of exposure factors, linearity and reproducibility were measured using an Unfors Xi dosimeter. The overall mean and range of ESAK during chest X-ray were 0.6 ± 0.3 (0.1-1.3) mGy, while for abdominal X-rays they were 4.0 ± 3.2 (1.3-9.2) mGy. Hospital with a CR system was found to use relatively higher doses. Dose values for abdominal X-ray procedures were comparable with previous studies. The dose for chest X-ray procedure was higher by a factor of 2-3 compared with the current international reference levels.

Original languageEnglish
Pages (from-to)102-106
Number of pages5
JournalRadiation Protection Dosimetry
Volume165
Issue number1-4
DOIs
StatePublished - 1 Jul 2015

Fingerprint

Dive into the research topics of 'Radiation dose levels for conventional chestand abdominal X-ray procedures in elected hospitals in Sudan'. Together they form a unique fingerprint.

Cite this